Personal Respiratory Protective Devices : Efficacy of Intranasal Stent with Filters
Tanpowpong K Mail

Intranasal, hollow, cylindrical, medical grade and silicone stent with two outer layers
face mask filters at both ends was proposed for atmospheric suspended particulate matter pre-
vention. The personal respiratory protective device efficacy was done at the Otolaryngology
Department, Ramathibodi Hospital from April 1996 to October 1997. Single pulse mode of carbon
dioxide laser smoke particle was the suitable source of atmospheric suspended particulate matter.
A laser plume evacuator removed laser smoke particles with 5 Millipore filters of 0.22 urn pore
size or 5 intranasal stent with filters attached at the inlet end. A Millipore filter got the same
laser smoke particle amount that passed through each intranasal stent filter with an air flow rate of
15 1/min controlled by a rotameter. Laser smoke particle deposition in filter materials was counted
under a high power optical microscope. Laser smoke particle amount in each layer of a four-
layer filter of intranasal stent with 7.5, 15.0 and 30.0 1/min air flow rates is shown. The filtration
efficacy of four, three and two layers of intranasal stent with a filter for laser smoke particle reten-
tion were compared. An intranasal stent with filter could be applied in a human nasal vestibule
with acceptable air flow resistant during public transportation in a traffic congested area.
Key word
: Intranasal, Silicone Stent, Filter, Efficacy, Suspended Particulate Matter, Laser
Smoke Particle