J Med Assoc Thai 2001; 84 (3):422

Views: 1,312 | Downloads: 17 | Responses: 0

PDF XML Respond to this article Print Alert & updates Request permissions Email to a friend


Comparative Study Between Westergren and Erythrocyte Sedimentation Pipette Method for Determination of Erythrocyte Sedimentation Rate
Siritantikom A Mail, Wiwanitkit V

A comparative study between standard Westergren and erythrocyte sedimentation pipette
(SediplastR) methods for determination of erythrocyte sedimentation rate was performed on 80
individual subjects. From the study, the comparison-of-methods plot SediplastR method (X)
vs
Westergren method (Y) gave the least square linear regression equation of Y
=
1.001 X -2.24
(r
=
0.99). Precision analysis gave a coefficient of variation below 3 per cent. The new erythro-
cyte sedimentation pipette method seems to be an effective and safe method for erythrocyte
sedimentation rate determination in the present day.
Key word : Erythrocyte Sedimentation Rate, Westergren Method, Pipette

Download: PDF