J Med Assoc Thai 2015; 98 (6):582

Views: 1,643 | Downloads: 85 | Responses: 0

PDF XML Respond to this article Print Alert & updates Request permissions Email to a friend


Decontamination Efficacy of Ultraviolet Radiation against Biofilms of Common Nosocomial Bacteria
Tingpej P Mail, Tiengtip R , Kondo S

Background: Ultraviolet radiation (UV) is commonly used to destroy microorganisms in the health-care environment. However, the efficacy of UV radiation against bacteria growing within biofilms has never been studied.

Objective: To measure the sterilization effectiveness of UV radiation against common healthcare associated pathogens growing within biofilms.

Material and Method: Staphylococcus aureus, Methicillin-resistant S. aureus (MRSA), Streptococcus epidermidis, Escherichia coli, ESBL-producing E. coli, Pseudomonas aeruginosa and Acinetobacter baumannii were cultivated in the Calgary Biofilm Device. Their biofilms were placed 50 cm from the UV lamp within the Biosafety Cabinet. Viability test, crystal violet assay and a scanning electron microscope were used to evaluate the germicidal efficacy.

Results: Within 5 minutes, UV radiation could kill S. aureus, MRSA, S. epidermidis, A. baumannii and ESBL-producing E. coli completely while it required 20 minutes and 30 minutes respectively to kill E. coli and P. aeruginosa. However, the amounts of biomass and the ultrastructure between UV-exposed biofilms and controls were not significantly different.

Conclusion: UV radiation is effective in inactivating nosocomial pathogens grown within biofilms, but not removing biofilms and EPS. The biofilm of P. aeruginosa was the most durable.

Keywords: Biofilms, Nosocomial infections, Ultraviolet radiation, Sterilization


Download: PDF