Radiation Dose to Medical Staff in Interventional Radiology
Oonsiri S , Jumpangern C , Sanghangthum T , Krisanachinda A , Suriyapee S Mail

Objective: The purposes of the present study were to determine the dose to medical staff in interventional radiology at different locations on the body measured by thermoluminescent dosimeter (TLD) and to relate the medical staff dose to patient dose measured by the dose-area product (DAP) meter.
Material and Method: The present study covered 42 patients in three interventional radiology procedures with three x-ray machines. Thermoluminescent dosimeters were stuck at eight positions on the radiologist’s skin during the procedure. In addition, direct reading from the DAP meter placed in front of the collimator of the x-ray tube, was recorded to estimate the patient radiation dose.
Results: The surface dose to the primary radiologist showed maximum value at the left forearm of 407 μGy. The ratios between the maximum interventional radiologist surface dose and patient dose are 12.88 μGy per 10 Gycm2 for transarterial oily chemoembolization TOCE (Siemens Polystar), 22.58 μGy per 10 Gycm2 for transarterial oily chemoembolization TOCE (Siemens Neurostar), 148.29 μGy per 10 Gycm2 for percutaneous transhepatic biliary drainage PTBD (Siemens Polystar) and 100.46 μGy per 10 Gycm2 for endoscopic retrograde cholangiopancreatography ERCP (GE Advantx).
Conclusion: The interventional radiologist surface dose can be estimated from the mentioned ratio if the patient dose is measured. This will help the radiologists to avoid receiving an excess dose during their work.

Keywords: Thermoluminescent dosimeter, Dose-area product, Radiation dose