Paskorn Sritipsukho MD*, Suthida Sritipsukho MD*, Pisit Wattanaruangkowit MD**
Affiliation : * Department of Pediatrics, Faculty of Medicine, Thammasat University, Bangkok, Thailand ** Department of Radiology, Faculty of Medicine, Thammasat University, Bangkok, Thailand
Objective : The present study aimed to validate the SSSL in pre-determinating the insertional length between
the sixth and tenth thoracic vertebrae (T6-T10) and examine repeatability property.
Material and Method: A prospective recruitment of 52 babies who were placed the umbilical arterial catheter
in the NICU, Thammasat University Hospital was performed. Insertional length of the distance between
suprasternal notch and superior iliac spine (SSSL) were performed. The catheter tips were verified against
anatomical points on chest and abdominal radiograph. The SSSL was measured twice on each patient to
indicate repeatability property.
Results : All babies (100 percent accuracy) were correctly placed the catheter tips at high placement (T6 to
T10) at the first attempt. Repeatability coefficient for repeating measure of the SSSL was 0.7 centimeters.
Conclusion : The SSSL is repeatable, simple, and perfectly accurate for pre-determination of the umbilical
arterial catheter length to position the catheter tip at T6 - T10.
Keywords : Umbilical arterial catheter, Simple, Accurate, Estimation, Length, High placement, Suprasternal notch, Superior iliac spine
JOURNAL OF THE MEDICAL ASSOCIATION OF THAILAND
4th Floor, Royal Golden Jubilee Building,
2 Soi Soonvijai, New Petchburi road,
Bangkok 10310, Thailand.
Phone: 0-2716-6102, 0-2716-6962
Fax: 0-2314-6305
Email: editor@jmatonline.com
» Online Submissions » Author Guidelines » Copyright Notice » Privacy Statement
» Journal Sponsorship » Site Map » About this Publishing System
© MEDICAL ASSOCIATION OF THAILAND. All Rights Reserved. The content of this site is intended for health professionals.