Submit manuscript

Accuracy and Clinical Utility of a Portable Coagulometer in an Emergency Setting

Chatchawan Rattanabannakit, MD*, Yongchai Nilanont, MD*, Chulaluk Komoltri, DrPH (Biostatistics)**, Naraporn Prayoonwiwat, MD*, Niphon Poungvarin MD, FRCP, FRI*

Affiliation : * Division of Neurology, Department of Medicine, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand ** Division of Clinical Epidemiology, Department of Research Development, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand

Background : There are limited data concerning accuracy of portable coagulometer in emergency setting.
Objective : To evaluate the accuracy of the CoaguChek® XS international normalized ratio (INR) monitor compared to the standard laboratory method in emergency department. Material and Method: Emergency room patients who required coagulation test were recruited. Parallel INR measurements between portable coagulometer and standard laboratory were performed.
Results : Fifty-five patients with a mean age of 59.1 + 15.7 years (20-87) were included. Men constituted 56.4%. Indications for testing were as follows: acute stroke 72.7%; abnormal bleeding 7.3%; taking anticoagulant 7.3%; and others 12.7%. Mean time + SD used from blood drawn to INR result report was 65.02 + 24.5 minutes for standard laboratory and 1 minute for portable coagulometer. Mean difference of INR result from portable coagulometer and standard laboratory was 0.02 + 0.13 and an excellence correlation between INR (r = 0.969) was demonstrated. There was no significant difference between the INR value from the two methods (p = 0.34).
Conclusion : The use of portable coagulometer (CoaguChek® XS) in emergency setting was accurate and required less time. Acute ischemic stroke patients are likely to benefit from a timely clinical decision making for thrombolysis medication.

Keywords : International normalized ratio (INR), Portable coagulometer, Point of care testing, Thrombolysis, Acute stroke


All Articles Download


INFORMATION

Contact info

JOURNAL OF THE MEDICAL ASSOCIATION OF THAILAND
4th Floor, Royal Golden Jubilee Building,
2 Soi Soonvijai, New Petchburi road,
Bangkok 10310, Thailand.
Phone: 0-2716-6102, 0-2716-6962
Fax: 0-2314-6305
Email: editor@jmatonline.com

JMed Assoc Thai
MEDICAL ASSOCIATION OF THAILAND
ISSN: 0125-2208 (Print),
ISSN: 2408-1981 (Online)
The content of this site is intended for health professionals.

Submissions

» Online Submissions » Author Guidelines » Copyright Notice » Privacy Statement

Other

» Journal Sponsorship » Site Map » About this Publishing System

© MEDICAL ASSOCIATION OF THAILAND. All Rights Reserved. The content of this site is intended for health professionals.