The Glucose Interference in Creatinine Measurement Using an Enzymatic Method: Effect of Creatinine Concentrations

Charoan Kaitwatcharachai MD*, Somthawil Kaitwatcharachai BNS**, Jaruda Aeden BSc ***, Dullached Wiriyasombat BSc***

Affiliation : * Division of Medicine, Hat-Yai Hospital, Songkhla, Thailand ** Division of Nursing, Hat-Yai Hospital, Songkhla, Thailand *** Division of Clinical Pathology, Hat-Yai Hospital, Songkhla, Thailand

Objective : To evaluate the influence of glucose and creatinine concentrations on the determination of creatinine by Jaffe picrate reaction and specific enzymatic assay. Material and Method: Unused dialysate with 4.25% dextrose was diluted to obtain seven glucose concentrations. Two series of dialysate were spiked with creatinine to yield concentrations of 5 and 10 mg/dl. Creatinine measurements were obtained by Jaffe method and enzymatic assay.
Results : In unused dialysate solution with glucose concentrations from 559 to 4,250 mg/dl, the creatinine values obtained by the Jaffe method were higher than the enzymatic assay (0.31 + 0.20 vs. 0.08 + 0.01 mg/dl, p < 0.05). The correlation coefficient between glucose and creatinine from the Jaffe method were 0.98 (p < 0.001) but showed no correlation with creatinine measured with the enzymatic assay. On the other hand, the mean values of creatinine in dialysate with creatinine concentrations of 5 and 10 mg/dl derived by Jaffe method were lower than enzymatic assay (5.74 + 0.12 vs. 6.16 + 0.36 mg/ dl and 11.56 + 0.17 vs. 12.69 + 0.66 mg/dl, respectively). At creatinine concentration of 10 mg/dl, the correlation between glucose concentration and creatinine from enzymatic assay was significant. In contrast, at creatinine concentration of 5 mg/ dl, the correlations obtained from both methods were significant.
Conclusion : The patterns of glucose interference with creatinine obtained from Jaffe method and enzymatic assay were quite different. The magnitude of interference with enzymatic assay was greater at a higher creatinine concentration. Therefore, the enzymatic assay might not be appropriate for creatinine measurement in patients using dialysate with dextrose 4.25% and membrane characteristic of high solute transporter.

Keywords : Creatinine measurement, Dialysate, Jaffe alkaline picrate reaction, Specific enzymatic assay


All Articles Download


INFORMATION

Contact info

JOURNAL OF THE MEDICAL ASSOCIATION OF THAILAND
4th Floor, Royal Golden Jubilee Building,
2 Soi Soonvijai, New Petchburi road,
Bangkok 10310, Thailand.
Phone: 0-2716-6102, 0-2716-6962
Fax: 0-2314-6305
Email: editor@jmatonline.com

JMed Assoc Thai
MEDICAL ASSOCIATION OF THAILAND
ISSN: 0125-2208 (Print),
ISSN: 2408-1981 (Online)
The content of this site is intended for health professionals.

Submissions

» Online Submissions » Author Guidelines » Copyright Notice » Privacy Statement

Other

» Journal Sponsorship » Site Map » About this Publishing System

© MEDICAL ASSOCIATION OF THAILAND. All Rights Reserved. The content of this site is intended for health professionals.