Comparison of Droplet Digital PCR and Real Time PCR Method for HBV DNA Quantification

Chatchawal Wongjitrat MD*, Navin Horthongkham PhD**, Ruengpung Sutthent MD, PhD**, Surangrat Srisurapanon PhD***

Affiliation : * Department of Internal Medicine, Faculty of Medicine, Srinakharinwirot University, Bangkok, Thailand ** Department of Microbiology, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand *** Department of Pathology, Faculty of Medicine, Srinakharinwirot University, Bangkok, Thailand

Background : HBV infection causes a potential serious public health problem. The ability to detect HBV DNA concentration is an important issue that had been continuously improved. When using quantitative polymerase chain reaction (qPCR), several factors are of concern, for example, sources of material, standard curve calibration, and PCR efficiency. Digital PCR (dPCR) is an alternative PCR-based technique for absolute quantification using Poisson’s statistics without requiring a standard curve.
Objective : Compare the data set of HBV DNA generated between dPCR and qPCR methods.
Materials and Methods : Fifty-four samples were quantified by Abbot’s real time PCR and with 2-6 log10 HBV DNA were selected for comparison with dPCR.
Results : Of these 54 samples, there were two outlier samples defined as negative by dPCR, whereas 52 samples were positive by both of these assays. The difference between two assays was less than 0.25 log IU/mL in 24/52 samples (46%) of paired samples; less than 0.5 log IU/mL in 46/52 samples (88%) and less than 1 log in 50/52 samples (96%). The correlation coefficient (r) was 0.788 (p-value <0.0001). Comparison with qPCR method, data generated by dPCR tend to be an overestimation in the sample with the low level of HBV DNA concentration and underestimated in the sample with high viral load. The variation of DNA by dPCR measurement might be due to the pre-amplification procedure and PCR template.
Conclusion : Measurement of HBV DNA by using dPCR, the results of the HBV DNA copy number tended to be deviated by over- or under-estimated when comparison to real time PCR method. In addition, a large quantity of DNA was used when compared to qPCR. However, the optimum processes of this assay have to be further investigated.

Keywords : Hepatitis B virus, Real time PCR, Digital PCR, DNA quantification


All Articles Download


INFORMATION

Contact info

JOURNAL OF THE MEDICAL ASSOCIATION OF THAILAND
4th Floor, Royal Golden Jubilee Building,
2 Soi Soonvijai, New Petchburi road,
Bangkok 10310, Thailand.
Phone: 0-2716-6102, 0-2716-6962
Fax: 0-2314-6305
Email: editor@jmatonline.com

JMed Assoc Thai
MEDICAL ASSOCIATION OF THAILAND
ISSN: 0125-2208 (Print),
ISSN: 2408-1981 (Online)
The content of this site is intended for health professionals.

Submissions

» Online Submissions » Author Guidelines » Copyright Notice » Privacy Statement

Other

» Journal Sponsorship » Site Map » About this Publishing System

© MEDICAL ASSOCIATION OF THAILAND. All Rights Reserved. The content of this site is intended for health professionals.